マルチメーター 3B SCIENTIFIC PHYSICS DT9201AのPDF インストラクションシートをオンラインで閲覧またはダウンロードできます。3B SCIENTIFIC PHYSICS DT9201A 5 ページ。 Digital multimeter

Operating temperature: 0° C – 40° C with
Storage temperature:
Electrical safety:
Safety specifications:
Excess voltage
category:
Contamination level:
Protection type:
Electromagnetic compatibility:
Interference emission: EN 55011:2009
Interference resistance: EN 61326-1:2013

Electrical specifications

V
Measuring range
200 mV
2 V, 20 V, 200 V
1000 V

DC voltage

Input impedance:
V
Measuring range
200 mV
2 V, 20 V, 200 V
750 V
AC voltage
Input impedance:
Frequency range:
A
Measuring range
20 µA
200 µA, 2 mA, 2 mA
20 mA, 200 mA
2 A, 20 A
DC current
Measuring voltage drop:
0 – 75% humidity
-10° C – 50° C with
0 – 75% humidity
EN 61010-1
CAT I: 1000 V
2
IP20
Accuracy
±0.5 % ± 3 digits
±0.8 % ± 2 digits
±1.0 % ± 2 digits
10 M
Accuracy
±1.2 % ± 5 digits
±1.0 % ± 5 digits
±1.2 % ± 5 digits
10 M
40 – 400 Hz
Accuracy
±1.8 % ± 2 digits
±2.0 % ± 2 digits
±2.0 % ± 10 digits
200 mV
A
Measuring range
20 µA, 200 µA, 2 mA
20 mA
200 mA
2 A, 20 A
AC current
Measuring voltage drop:
Frequency range:
Measuring range
200
2 k , 20 K , 200 k
2 M
20 M
Accuracy is given for 1 year after calibration at
23° C ±5° C, RH<75%.

5. Description

Digital multimeter E is a robust, battery operated
multimeter with a 3½-digit LCD display for
measuring voltage, current and resistance as
well as for diode and hFE gain testing.
All measurement ranges are selected by means
of a rotary dial. All measurement ranges are
protected against overload except the 20 A
range.
The meter is equipped with a hold function,
negative polarity indication, over range indica-
tion, low battery indication and automatic switch
off after 15 minutes. After the power is automati-
cally switched off it needs to be turned off and
turned on again to continue the power.
The digital display is folding for ease of reading
and on the backside there is a fold out prop for
standing the device on a table.

6. Operation

Note
The multimeter switches into an indeterminate
state if the Hold button is activated (Hold func-
tion button (8) pressed). After it is switched on,
the display will then show "1.666" (where the
position of the decimal point is dependent on the
selected measuring range). If this case should
arise, deactivate the Hold function while the
equipment is switched on (deactivate Hold func-
tion by pressing the button one more time).
3
Accuracy
±2.0 % ± 3 digits
±2.0 % ± 5 digits
±2.5 % ± 10 digits
200 mV
40 – 400 Hz
Accuracy
±1.0 % ± 10 digits
±1.0 % ± 4 digits
±1.0 % ± 10 digits