2. Installation

2-2. Running the MED64-Allegro System and checking the noise

Install the Test board to the MED Multi-well Connector and check the noise. The baseline noise should
be around several microvolts at all 64 channels as seen in the Figure 6. If noise is larger than a few
microvolts, check the followings:
1. All equipment are grounded correctly.
2. Amplifiers, MED Multi-well Connector, and all Connector cables are placed AWAY from any equipment
which generates electric or magnetic field. (e.g. power supply unit, peristaltic pump, heater). Pay
particular attention for the amplifiers' power supply unit.
3. All cables are firmly connected.
Figure 6. Appropriate noise level for the MED64-Allegro System