顕微鏡 Olympus BX2 SERIESのPDF メンテナンス・マニュアルをオンラインで閲覧またはダウンロードできます。Olympus BX2 SERIES 29 ページ。 Olympus microscope bx2
Olympus BX2 SERIES にも: インフォメーション (6 ページ)

Olympus BX2 SERIES メンテナンス・マニュアル
BX2 MAINTENANCE MANUAL

1. Optical adjustment

PREPARATION
If the left/right optical axis is remarkably displaced at
checking, perform the following adjustment.
Align the specimen center with the cross center of the
WH10X eyepiece by turning the control knob of the stage.
Before adjustment
Cross center of
cross WH10X
REPAIR PROCEDURE
Chapter 3.
REPAIR PROCEDURE
Adjusting the left/right optical axis
*Insert the cross eyepiece
into the right sleeve.
specimen
whose center is
identified
(Concentric circles etc.)
specimen center
*Adjust the interpupillary
distance to about 62mm
(See the illustration).
position of
approx. 62mm
After adjustment
- 11 -
60
50