Olympus BX2 SERIES メンテナンス・マニュアル
BX2 MAINTENANCE MANUAL
ADJUSTMENT THE LEFT/RIGHT OPTICAL AXIS
(1) Moving the cross eyepiece to the left sleeve.
Work
Move the cross eyepiece to the left sleeve.
(2) Aligning the cross center of eyepiece with the specimen center.
Work
1. Loosen the two screws slightly which
secure the left sleeve.
2. Align the center of eyepiece with the
specimen center while observing through
the cross WH10X.
(Change the left sleeve position by
hand.)
REPAIR PROCEDURE
If the optical axis between left and right sleeve
is deviated, the center of the specimen and the
cross center of eyepiece are also deviated.
60
50
Image at first
60
50
The center are aligned.
60
50
- 12 -
Image seen through the cross eyepiece
Image seen through the cross eyepiece