Use only a PE Photo eyepiece for photomicrography.
Insert the Photo
eyepiece @ with the desired magnification into the straight Photo tube
mounted on the trinocular Observation tube.
(Fig. 43)
Place the Camera unit directly over the circular dovetail of the straight
Photo tube. Make sure the index dots @ on the straight Photo tube and
the Camera unit are aligned, then clamp the unit.
5
2
f.2
See page 15 of the "Observation
Tube" section.
1. Camera focusing is done using the binocular section of the trinocular
Observation tube.
+ Whenever
you remove the focusing telescope from the photo-
graphic unit, be sure to install the dust cap.
2. Insert a finder eyepiece into the right eyepiece sleeve.
3. The finder eyepiece has a reticle with four masks and a double crossline,
and the focus is for all purposes the same for the reticle and the Camera
film plane. The masks indicate the areas covered, and the numerals next
to the masks correspond to the magnification of the Photo eyepiece.
Different finder eyepieces are available for different cameras. Select the
type that is appropriate for the Camera being used.
4. Because of the great depth of focus of 1.25X to 4X objectives, use of a
focusing magnifier (U-FT) is recommended for accurate focusing.
0 When using low power objectives, focusing may be accomplished easier
when using the focusing telescope on the Camera unit rather than the
finder eyepiece.