- ページ 5

顕微鏡 Olympus BX43のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Olympus BX43 48 ページ。 System microscope

Olympus BX43 取扱説明書

IMPORTANT

SAFETY PRECAUTIONS
1
Fig. 1
1. After the equipment has been used in an observation of a specimen that
is accompanied with a potential of infection, clean the parts coming in
contact with the specimen to prevent infection.
· Moving this product is accompanied with the risk of dropping the
specimen. Be sure to remove the specimen before moving this
product.
· In case the specimen is damaged by erroneous operation, promptly
take the infection prevention measures.
· Follow the procedures described in Chapter "Getting Ready" (see on
page 2) prior to using the accessories of this product. Otherwise, the
stability of the device will be lost and the dropped specimen will cause
the possibility of infection.
2. Install the microscope on a sturdy, level table or bench.
3. Be careful not to tilt the microscope too much. Otherwise, the mobile
parts such as the light path selector knob may move in unintended
directions.
4. Always use the power cord provided by Olympus. If no power cord is
provided, please select the proper power cord by referring to the section
"PROPER SELECTION OF THE POWER SUPPLY CORD" at the end of
this instruction manual.
If the proper power cord is not used, product safety performance cannot
be warranted.
5. Always ensure that the grounding terminal of the microscope and that of the
wall outlet are properly connected. If the equipment is not grounded, Olympus
can no longer warrant the electrical safety performance of the equipment.
Operation Using the LED Lamp
Safety Note on LED (Light Emitting Diode)
The LED incorporated in the LED lamp housing is a class 1 LED
product as defined in the applicable standard. The light emitted by LED
is basically safe, but do not look at the illumination light directly for an
extended period to prevent your eye from being injured
Operation Using the Halogen Bulb
1. To avoid potential shock hazards and burns when replacing the halogen
bulb, set the main switch of the TL4 power supply unit to "
allow the lamp socket @ and bulb to cool before touching them.
Designated bulb
Always use the designated lamp bulb. Using other bulb may
CAUTION
cause malfunction.
2. If the lamp cable or power cord comes in contact with the lamp socket
or its surroundings, the cable or cord may be melt by the heat, causing
an electric shock or fire hazard. To prevent this, distribute the lamp cable
and power cord at enough distances from the lamp socket.
3. The standard service life of the lamp housing is eight (8) years of use or
20,000 hours of total power ON period, whichever is the shorter period.
For details, see Inspection Sheet on page 38.
CLASS 1 LED PRODUCT
6V30WHAL (PHILIPS 5761)
BX43
(Fig. 1)
" (OFF) and
1