- ページ 17

顕微鏡 Olympus BX51のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Olympus BX51 34 ページ。 System
Olympus BX51 にも: 取扱説明書 (17 ページ), 取扱説明書 (40 ページ), 取扱説明書 (37 ページ), 取扱説明書 (11 ページ)

Olympus BX51 取扱説明書

4-2 Intermediate Polarizing Attachment

@
Fig. 17
Fig. 18
15
1
Using the Bertrand Lens (U-CPA only)
By manipulating the Bertrand lens knob @ on the front, the Bertrand
lens is inserted into or removed from the light path. At the pushed-in
({) position, the lens is engaged. At the pulled-out (¦) position, the
lens is removed from the light path.
2
Using the Analyzer
The analyzer @ is engaged when pushed in to the second click-stop
position. To disengaged the analyzer, and to place the empty hole in
the light path, pull the analyzer out to the first click-stop.
3
Focusing the Conoscopic Image (U-CPA only)
To focus the conoscopic image, manipulate the Bertrand lens focusing
slider ². While observing the conoscopic image, move the slider to
find the position where image is sharply focused.
²
@
(Fig. 17)
(Fig. 18)
(Fig. 18)