CX21 MAINTENANCE MANUAL
5. Replacing Pinion Ass'y of Plane Stage
Grease;
OT2008
(D)
Left and right
dovetails
Loosen the stopper
until it removes from
the groove of
condenser.
Grease;
OT2008
sliding
surface
(E)
(1) Loosen the stopper (A) and remove the condenser (B) downward by turning
the knob (F).
(2) Remove the left dovetail (C) as seen from the front side.
Screws : AB3X8SA, 2pcs. (*1)
(3) Remove the right dovetail (D).
Screws : AB3X8SA, 2pcs. (*2)
(4) Remove the pinion spring (E).
(5) Remove the pinion ass'y (F).
(6) Assemble the reverse order of disassembly.
Note on assembly :
1) Apply grease to the portions shown as the above figure.
2) The right dovetail (D) is mounted by pushing it in the arrow directions.
3) The left dovetail (C) is mounted by pushing it in the upward direction.
At this time, adjust the position of dovetail (C) in the left and right directions so that
the condenser moves smoothly without a play (vertical movement).
*2
(A)
Grease;
OT2008
(B)
Front side
*1
(C)
(F)
A009
sliding
surface
- 19 -
REPAIR PROCEDURE
: Pushing direction
: Adjust the position of
dovetail (C).