- ページ 16

顕微鏡 Olympus CX41のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Olympus CX41 32 ページ。 System microscope
Olympus CX41 にも: 概要 (7 ページ), インストラクション (32 ページ)

Olympus CX41 取扱説明書
Fig. 17
Fig. 18
Fig. 19
CH2-DS
Fig. 20
13
2

Aperture Iris Diaphragm

· The aperture iris diaphragm determines the numerical aperture of the
illumination system. Matching the numerical aperture @ of the
illumination system with that of the objective provides better image
resolution and contrast, and also increases the depth of focus.
· Since the contrast of microscope specimens is ordinarily low, setting
the condenser aperture iris diaphragm to between 70% and 80% of
the N.A. of the objective in use is usually recommended. If necessary,
adjust the ratio by removing the eyepieces and looking into the
eyepiece sleeves while adjusting the aperture iris diaphragm knob ²
until the image shown in Fig. 17 is seen. (Fig. 18)
#If the aperture iris diaphragm is set too small, image ghost may be
observed.
Using the numerical aperture scale:
}Align the aperture iris diaphragm knob ² with the NA value @ on the
scale. The scale value should correspond to the NA number engraved
on the objective.
Since the aperture iris diaphragm knob has a certain width, align the
center line engraved on the knob with the scale indication.
When using a 100X objective, turn the aperture iris diaphragm knob ²
all the way to 0.9 on the scale. (Fig. 18)
3

Using Filters

}You can place a filter in the light path with either method.
· Insert a filter with diameter of 32.5 mm @ in the condenser accessory
(CH2-FH or CX-AL) attached below the condenser.
· Insert one or more filter with a diameter of 45 mm ² on the light exit
glass on the microscope base.
}For the types of the filters, please consult Olympus or its catalogues.
4

Using Darkfield Ring CH2-DS

}The CH2-DS can be attached in the same way as inserting the 32.5
mm filter.
Usage
Inserting the darkfield lens makes possible darkfield observations
using a 4X to 40X objective.
(Figs. 17 & 18)
(Fig. 16)
(Fig. 19)
(Fig. 20)