3-7 Simplified Phase Contrast Ring Slits CX-PH1/PH2/PH3

1
Appearance
Ring Slits
CX-PH1/PH2/PH3
Phase Contrast Objectives
PlanCN-Ph series (10X, 20X, 40X, 100XO)
2
Installation
Attach a ring slit in the same way as a filter holder.
#Be sure to replace the objective with a phase contrast objective.
Fig. 23
Fig. 24
3
Operation
1. Turn the revolving nose piece to engage in light path the phase
contrast objective lens with the same value as the ring slit in use.
2. Place the specimen and bring it in approximate focus.
3. Remove the eyepieces and replace them with the CT-5 centering
telescope.
4. Turn the upper ring of the CT-5 to adjust the focus so that the bright
ring (ring slit) @ and dark ring (objective's phase plate) ² are seen
clearly in the field of view. (Fig. 23)
5. Rotate the two centering knobs ³ so that the bright and dark rings
overlap concentrically. (Figs. 23 & 24)
6. Remove the CT-5, replace it with the eyepieces and start phase
contrast observation.
7. To increase the image contrast, place the green filter (45 mm diam.) on
the light exist glass of the microscope.
CX41
Green Filter
45G533 or 45IF550
Centering Telescope CT-5
(Figs. 23 & 24)