2-2 Detailed Assembly Procedures
²
³
@
|
ƒ
Fig. 1
†
...
Fig. 2
1
Connecting the Cable and Power Cord
The lamp housing cable and power cord are vulnerable to bend or
twist. Do not apply excessive force to them.
Be sure to set the main switch to "
lamp housing cable and power cord.
1. Insert the cable @ from an applicable lamp housing into the lamp housing
connector ² by aligning the connectors, and secure using the clamping
ring ³ on the lamp housing connector.
Always use the power cord provided by Olympus. If no power cord
is provided, please select the proper power cord by referring to the
section "PROPER SELECTION OF THE POWER SUPPLY CORD" at
the end of this instruction manual.
2. Insert connector | of the power cord into connector ƒ of the power
supply firmly.
Always use the power cord supplied by Olympus and connect it to a
3-conductor wall outlet with a grounding terminal. If the grounding
terminal of the wall power outlet is not grounded/earthed, Olympus
can no longer warrant the electrical safety performance of the
equipment.
3. Insert the plug ... of the power cord into the wall power outlet †.
U-RFL-T/U-RX-T
(Figs. 1 & 2)
" (OFF) before connecting the
4