- ページ 25

医療機器 Olympus UHI-2のPDF メンテナンス・マニュアルをオンラインで閲覧またはダウンロードできます。Olympus UHI-2 38 ページ。

Olympus UHI-2 メンテナンス・マニュアル

3-3 Rear panel unit

3-4 Power supply unit

CCUK3x6SA (Nejilock coated)
CCUK3x6SA
DISASSEMBLING/ASSEMBLING PROCEDURE
Disassembling
1.
Remove the harness from the rear panel
unit.
2. Remove the screw HCBK4x8SA (3 pcs.)
which secure the rear panel.
3. Remove the rear panel unit from the main
body.
Philips screwdriver No. 2
Assembling
1. Mount the rear panel unit on the main
body.
2.
Secure the rear panel unit to the main body
with the screw HCBK4x8SA (3 pcs.).
3. Connect the harness to the rear panel unit.
Philips screwdriver No. 2
Disassembling
1.
Remove the screw CCUK3x6SA (4 pcs.)
which secure each converter.
Philips screwdriver No. 2
Assembling
1
Secure the each converter with the screw
CCUK3x6SA (4 pcs.).
For the location shown on the left,
apply Nejilock to the screw thread
before tightening.
Nejilock
Philips screwdriver No. 2
3-4