3-5 K-connector unit

CCUK4x6SZ
st
3-6 1
pressure reducing unit
CCUK4x8SZ
Connect to the
electric pneumatic
unit.
S-ring
DISASSEMBLING/ASSEMBLING PROCEDURE
Connect to the
UPUH15U J2.
S-ring
3-5
Disassembling
1.
Remove the S-ring from the K-connector
with a spanner.
2.
Remove the screw CCUK4x6SZ (4 pcs.)
which secure the K-connector unit.
Assembling
1.
Secure the K-connector unit with the screw
CCUK4x6SZ (4 pcs.).
2.
Fix the S-ring temporarily to the K-
connector unit.
Fix the S-ring after mounting the
electric pneumatic ratio valve.
3.
Connect the harness to the UPUH15U J2.
Spanner
Philips screwdriver No. 2
Disassembling
1.
Remove the S-ring from the 1st pressure
reducing unit with a spanner.
2.
Remove the screw CCUK4x8SZ (4 pcs.)
which secure the 1st pressure reducing
unit.
Spanner
Philips screwdriver No. 2
Assembling
1.
Secure the 1st pressure reducing unit with
the screw CCUK4x8SZ (4 pcs.).
2.
Fix the S-ring temporarily to the 1st
pressure reducing unit.
3.
Fix the S-ring completely to the 1st
pressure reducing unit.
4.
Connect the tube from the electric
pneumatic unit.
Spanner
Philips screwdriver No. 2