- ページ 17
試験装置 AccuPulse NIBP SimulatorのPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。AccuPulse NIBP Simulator 20 ページ。 Hand held nibp simulator
Using Commands
The remote-control user of an ApHandheld utilizes the following command to enable
remote-control:
<> <CR>
In response to this, the user should see the message:
<USB_MULTICHAR_CMDS_ENABLED>
=
>
The first line returned is a comment to the user "<USB...>" which indicates the
command parser is awake.
The '=' indicates the command was recognized and hence was executed successfully.
The '>' is the prompt char, indicating the unit is waiting for another command.
Argument versus Argument-less commands
Commands that take arguments give their arguments Carriage-return delimited from
the command. Where this documentation shows arguments in braces
[argument1,argument2] it is listing acceptable variables. These are to be entered
without the braces.
Sent:
< ARGUMENT_COMMAND > <CR>
ARGUMENT1 <CR>
Received:
= <CR> --OR--
>
Sent:
< ARGUMENT_LESS_COMMAND > <CR>
Received:
= <CR> --OR-
>
Note: Synch on the '>' character (prompt character) to indicate a
command completed. The previous line in the stream will have the result
of the operation ('=', '*', etc).
Commands
Start Remote Control:
Send:
<> <CR>
Recv:
<USB_MULTICHAR_COMMS_STARTED> <CR>
= <CR>
>
Keypad Disable / Enable Commands
Send: <KEYPAD_OFF> <CR> --OR--
Recv: '=' <CR>
Use of this is highly recommended. These commands turn processing of the physical
keypad off/on so that the user does not have to worry about keypad inputs interfering
with the execution of a remote control session.
(error code if error occurred) <CR>
(error code if error occurred) <CR>
<KEYPAD_ON> <CR>
Help
This command returns a listing of available commands in the current software.
These may vary based upon the configuration options loaded in the unit.
Send:
<HELP>
Recv:
Available Remote Commands:
<TEST_MODE>
<SET_SHFT>
<SET_PRESET>
<SET_PR>
<SET_DEVICE_ID>
<SET_AMP>
<SETP_CFG>
<MAN_STR>
<MAN_SMP>
<LEAK_CFG>
<KEYPAD_ON>
<KEYPAD_OFF>
<I>
<HEX_SCREENSHOT>
<HELP>
<GET_SERIAL_NUM>
<GET_FEATURES>
<GET_DEVICE_ID>
<GET_CALTABLE>
<GET_BP_PARAMS>
<BUTTON>
<BIN_SCREENSHOT>
=
>
Virtual Button Press
This provides a way to push the front panel buttons by remote control. All operations can
always be controlled in this way. In some cases a direct remote control command is also
provided as a shortcut to perform a certain task.
Send:
<BUTTON> <CR>
KEYNAME <CR>
Recv: '=' – Success
'*' – key blocked this mode (keypad is off)
'#' – bad key name
Key Names:
All handhelds: F1,F2,F3,F4,UP ,DWN,LEFT,RIGHT,MENU
Generic Keypad: ESC, ENTER, SCROLL_UP , SCROLL_DOWN
Fixed Keypad: BP ,LEAK,OVERP ,METER
Selecting Test Mode
Set Test Mode
Send: <TEST_MODE> <CR>
[BP , LEAK, OVERP , SETP] <CR>
Recv: '=' – success
Setting Test Parameters
Set BP Mode
BP_MODE is the default "Preset" mode included in all NIBP units.
UTP_MODE is the optional User Defined Test Points software module.
Send:
<SET_BP_MODE> <CR>
[BP_PRESET,UTP_MODE] <CR>
Recv:
'=' – success
'#' – invalid mode name.
17