- ページ 5
産業機器 Siemens SIREMOBIL Iso-CのPDF 設置・設定マニュアルをオンラインで閲覧またはダウンロードできます。Siemens SIREMOBIL Iso-C 6 ページ。
Siemens SIREMOBIL Iso-C にも: 取り付けとセッティングの説明 (4 ページ), 取扱説明書 (16 ページ), メンテナンス方法 (46 ページ), 取付説明書 (6 ページ)
Installing a Video Printer
•
Switch the SIREMOBIL Iso-C off and on again to activate the new configuration.
For Memoskop C-E (without keyboard)
•
Switch on the system.
•
Connect the service PC to the service interface of the SIREMOBIL Iso-C.
•
Install and start-up the MEMOSKOP service program on the service PC.
•
Program "Hardcopy release" on SIREMOBL Iso-C systems with one monitor.
•
Program "Hardcopy release" on SIREMOBIL Iso-C systems with two monitors.
•
If there is an area dose product measurement device present (Diamentor), activate
"Dose on hardcopy output" as well.
•
Transfer the parameters to the MEMOSKOP.
•
Exit the MEMOSKOP service program.
Function test
•
Call up a fluoroscopic image on monitor A or monitor B.
•
Release an exposure at the operating panel of the basic system and at the monitor
trolley keyboard.
Final steps
•
Switch off the SIREMOBIL Iso-C.
•
Test the protective conductor using the protective conductor meter 44 15 899 RV 090.
Refer to ARTD - 002.731.17...
Changes as compared to previous versions
•
Paragraph on page 3 and 4 "Power cable" changed.
TD SD 24 / Arnold
Siemens AG
Medical Engineering
The video printer should print an image in both cases.
Register 3
TD SD 24 / Groß
SPR2-230.033.02
Rev. 02
10.99
SMS Iselin / O'Donnell
Page 5 of 6
SIREMOBIL Iso-C
TD SD 24
5
1
1
1