6.4

20 mA calibration

6.5

Blanking

Blanking is used to ignore the zone in front of the transducer where false echoes are at a level
that interfere with the processing of the true echo. It is measured outward from the sensor
face. The minimum recommended blanking value is 0.25 m (0.82 ft) but can be increased in
order to extend the blanking.
SITRANS LU180
Operating Instructions, 09/2021, A5E37100674-AB
20 mA calibration initiated
view stored 20 mA calibration value
(i.e. 0.50 m)
press "4" to increase to new calibration value
(i.e. 0.45 m)
new calibration value