Connecting
4.1
Safety instructions
4.1.1

Laws and directives

Observe the safety rules, provisions and laws applicable in your country during connection, assembly and operation. These
include, for example:
● National Electrical Code (NEC - NFPA 70) (USA)
● Canadian Electrical Code (CEC) (Canada)
Further provisions for hazardous area applications are for example:
● IEC 60079-14 (international)
● EN 60079-14 (EU)
WARNING
Hazardous contact voltage
Risk of electric shock in case of incorrect connection.
For the electrical connection specifications, refer to the information in Electrical connections (Page 19).
At the mounting location of the device observe the applicable directives and laws for installation of electrical power
installations with rated voltages below 1000 V.
WARNING
Missing PE/ground connection
Risk of electric shock.
Depending on the device version, connect the power supply as follows:
Power plug: Ensure that the used socket has a PE/ground conductor connection. Check that the PE/ground conductor
connection of the socket and power plug match each other.
Connecting terminals: Connect the terminals according to the terminal connection diagram. First connect the PE/ground
conductor.
Note
Electrical circuit breaker in accordance with IEC 60947-1 and IEC 60947-3
In accordance with IEC 60947-1 "Standard for low-voltage switchgear and controlgear" and IEC 60947-3 "Switches,
disconnectors, fuses" you require an electrical circuit breaker for the device.
We recommend commercially available automatic circuit breakers.
4.1.2
Analyzers in hazardous areas
DANGER
Explosion hazard
If a flammable or ignitable atmosphere exists, plugs must never be disconnected or lamps/fuses replaced when the
analyzer is supplied with power.
DANGER
Potentially explosive atmosphere
The analyzer keys must not be pressed if a potentially explosive atmosphere may be present.
If operation using the keyboard is necessary, a hot work permit is absolutely essential.
Exceptions: Analyzers with approvals 1708 or 2022
16
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
A5E45779389002-01, 09/2018