4
Connecting
4.1
Safety instructions
4.1.1
Laws and directives
Observe the safety rules, provisions and laws applicable in your country during connection, assembly and operation. These
include, for example:
● National Electrical Code (NEC - NFPA 70) (USA)
● Canadian Electrical Code (CEC) (Canada)
Further provisions for hazardous area applications are for example:
● IEC 60079-14 (international)
● EN 60079-14 (EU)
WARNING
Hazardous contact voltage
Risk of electric shock in case of incorrect connection.
For the electrical connection specifications, refer to the information in Electrical connections (Page 19).
•
At the mounting location of the device observe the applicable directives and laws for installation of electrical power
•
installations with rated voltages below 1000 V.
WARNING
Missing PE/ground connection
Risk of electric shock.
Depending on the device version, connect the power supply as follows:
Power plug: Ensure that the used socket has a PE/ground conductor connection. Check that the PE/ground conductor
•
connection of the socket and power plug match each other.
Connecting terminals: Connect the terminals according to the terminal connection diagram. First connect the PE/ground
•
conductor.
Note
Electrical circuit breaker in accordance with IEC 60947-1 and IEC 60947-3
In accordance with IEC 60947-1 "Standard for low-voltage switchgear and controlgear" and IEC 60947-3 "Switches,
disconnectors, fuses" you require an electrical circuit breaker for the device.
We recommend commercially available automatic circuit breakers.
4.1.2
Analyzers in hazardous areas
DANGER
Explosion hazard
If a flammable or ignitable atmosphere exists, plugs must never be disconnected or lamps/fuses replaced when the
analyzer is supplied with power.
DANGER
Potentially explosive atmosphere
The analyzer keys must not be pressed if a potentially explosive atmosphere may be present.
If operation using the keyboard is necessary, a hot work permit is absolutely essential.
Exceptions: Analyzers with approvals 1708 or 2022
•
16
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0