Inflowing medium
Product flow
half a turn. This is sufficient to reach the recommended installation
position.
In adhesive and viscous media, the surfaces of the tuning fork should
protrude into the vessel to avoid buildup. Therefore nozzles for flanges
and mountings bosses should not exceed a certain length.
Fig. 6: Adhesive products
If SITRANS LVL100 is mounted in the filling stream, unwanted false
measurement signals can be generated. For this reason, mount
SITRANS LVL100 at a position in the vessel where no disturbances,
e.g. from filling openings, agitators, etc., can occur.
To make sure the tuning fork of SITRANS LVL100 generates as little
resistance as possible to product flow, mount the sensor so that the
surfaces are parallel to the product movement.