- ページ 2

試験装置 Alcolizer Easy CheckのPDF クイック・スタート・マニュアルをオンラインで閲覧またはダウンロードできます。Alcolizer Easy Check 2 ページ。

Getting the most out of your Easy Check
• It is recommended to not consume
alcohol, eat, use mouthwash or breath
spray for 15 minutes prior to testing.
• Breath Alcohol Concentration (BAC)
can rise for up to two hours after
last drink. Re-test if in doubt.
• It may take 10 hours or more for
a high BAC level to return to zero.
Further testing later in the day or
the next morning is recommended.
• Calibration is required after six
months. Easy Check will display
'RECAL'. You cannot use the
breathalyser until it is calibrated.
• Do not store unit in places where
moisture or extreme temperature
conditions can occur.
Power Usage/ Batteries
Easy Check is powered by x2 Alkaline
batteries (AAA). Do not replace batteries
- this is done during calibration. Removal
of batteries will incur loss of calibration
and warranty. Note: When not in use,
the back light will turn off and the device
will go into low power standby mode.
Designed, developed and manufactured in Australia by
Alcolizer Technology – the largest supplier of alcohol
detection equipment to Australian law enforcement.
Steps to conduct an Easy Check test:
1. Fit a mouthpiece onto the inlet located
on the (front) left side of the device.
2. Ensure the top-left of the screen
displays 'READY'– you can then press
the button to conduct a test.
3. Easy Check will display an animation
whilst it conducts a quick self-check.
4. When the display shows 'blo', take a
deep breath, and blow steadily into the
mouthpiece. Short beeps will sound
while blowing. Continue to blow until a
click and short musical tone is heard.
5. After a few seconds, your result will be
displayed – Example: '.030g/100ml'.
Note: If you do not blow within 15
seconds, or hard or long enough
a musical tone will sound and the
message, 'TRY AGAIN' will be displayed.
Repeat step 2 onwards to try again.
Recovery period is usually less than one
minute but recovery may take longer after a
high BAC reading.
Easy Check will be ready for a new test
when its recovery period is complete and
the 'READY' message is displayed.
Drive
safe.
Drive
confident.
1300 789 908
|
www.easycheck.com.au
Excessive Readings
The display will flash if your reading
exceeds the device limit of .300g/100ml.
Other Functions
Press and hold the button to display the
following information:
• The last reading
• The number of tests remaining
till calibration
• The number of days before
calibration is due
The information will continue to be
displayed, in sequence for three seconds
each, for as long as the button is
pressed.
Display Illumination
The display light automatically turns off
after a few seconds.
If the result is displaying but 'READY' is
not yet displayed, turn the display light
on again by briefly pressing the button.
Note: When the Easy Check is in low
power standby mode (not backlit), it
will display 'READY' at the top-left of the
screen and also the last BAC test result.
V2