P. 4
It is a state of the art instrument used to detect Ethyl Alcohol from a valid breath
sample. It uses an electrochemical fuel cell to measure the concentration of
alcohol from the sample of the expired breath from a human subject. If alcohol
is present, a corresponding voltage is generated from the fuel cell, which is
proportional to the alcohol content of the sample provided by a human subject.
This voltage is then sent to the CPU and converted to a subject's BAC which is
then displayed.
Instrument Features:
1. 65,000 test records can be stored and uploading to the PC for record
keeping.
2. Use disposable mouthpieces or sampling cups to prevent foreign matter from
entering.
3. A hard plastic case is included for storing the device and accessories for easy
transport. When not in use, the device and accessories should be stored in
this plastic case to prevent damage.
4. Ability to run as often as needed. Recovery time between tests on average
is less than 10 seconds between positive tests.
5. Powered by High-capacity lithium battery.
sliding cover on the back of the device. The level of power left in the battery
is indicated on the screen. A cigarette plug adapter is also included for use
by the operator.
Storage temperature range is -25°C to 70°C, storage humidity is 15% to 90%,
6.
inappropriate storage of the device will shorten the lifespan of the fuel cell
sensor.
INTRODUCTION
Battery is located behind the