PD030 DIODE LASER MANUAL
4. Operation Instructions continued.
c. Output time setting
There three ways of output:
1.
Free output: The laser emits when footswitch is pressed, it is also called real-time
output.
2.
Pre-set pulse amount: Input the amount of pulses need to output. When the
footswitch is pressed, the laser emits pulse of the amount you set.
3.
Pre-set time: Estimate and then input the treatment time. When the footswitch is
pressed, the laser emits for the time you set.
Caution: Pre-set pulse and Pre-set time are designed for experienced users.
We strongly recommend that new users use free output.
Press the output mode you want and set the value(same as detail parameter setting), then
press "Enter" to confirm.
4.4 Change Status and Emit
After all parameters are set and confirmed, press "Ready" to switch to ready status, the
parameters will be locked.
Then press the footswitch, laser will emit as you set.
If you want to change parameters, press "Stand by" to switch to stand-by status.