5. Service Instructions
5.1 Serviceable parts
With necessary tools and skills, the following parts are replaceable:
Rubber pad
1.
Needle
2.
Spring stop
3.
Leaf spring
4.
Note: Before service of any kind, All parts must be cleaned according to the cleaning
instructions.
Warning: The sharp needle will be exposed during removal. It presents a potential hazard
from possible contact with infected blood samples. Take all necessary precaution when
handling de-capper head or chute.
1. Replacement of Rubber Pad on Tube Support
When the Tube support no longer holds the Tubes in a steady grip at de-capping, the
Rubber pad should be replaced.
Tube support
without pad
Remove the old Rubber pad from the Tube support.
•
Clean the Tube support with Acetone.
•
Remove protective film from the rubber pad and put it onto the tube support.
•
Press firmly on the rubber pad.
•
[email protected] · 965 116 521 · 902 222 275 · www.akralab.es
Rubber pad with self-adhesive tejp
Figure 6: Tube support rubber pad