5. Service Instructions

5.1 Serviceable parts
With necessary tools and skills, the following parts are replaceable:
Rubber pad
1.
Needle
2.
Spring stop
3.
Leaf spring
4.
Note: Before service of any kind, All parts must be cleaned according to the cleaning
instructions.
Warning: The sharp needle will be exposed during removal. It presents a potential hazard
from possible contact with infected blood samples. Take all necessary precaution when
handling de-capper head or chute.
1. Replacement of Rubber Pad on Tube Support
When the Tube support no longer holds the Tubes in a steady grip at de-capping, the
Rubber pad should be replaced.
Tube support
without pad
Remove the old Rubber pad from the Tube support.
Clean the Tube support with Acetone.
Remove protective film from the rubber pad and put it onto the tube support.
Press firmly on the rubber pad.
[email protected] · 965 116 521 · 902 222 275 · www.akralab.es
Rubber pad with self-adhesive tejp
Figure 6: Tube support rubber pad
10