5. MAINTENANCE

5.1 General Maintenance
Periodically wipe the case with a damp cloth and mild
detergent. Do not use abrasives or solvents.
Dirt or moisture in the terminals can affect readings.
To clean the terminals:
- Turn the meter off and remove all test leads.
- Shake out any dirt that may be in the terminals.
- Soak a new swab with a cleaning and oiling agent.
- Work the swab around in each terminal. The oiling agent
insulates the terminals from moisture-related contamination.
To avoid electrical shock or damage to the
meter, do not get water inside the case.
Remove the test leads and any input signals
before opening the case
5.2 Fuse replacement
To replace the Meter's fuse (see the Figure):
- Set the rotary switch to OFF range.
- Disconnect test leads from any inputs terminals.
- Use a screwdriver to unscrew the two screws secured on
the battery cover.
- Separate the battery cover from the bottom case.
- Replace the fuse only with specified ratings:
F 200mA/250V ∅5×20 (quick acting).
- Replace the battery cover and secure by the two screws.
Before replacing the fuse, disconnect test
leads from any circuit under test.
To prevent damage or injury, replace the fuse
only with specified ratings.