5-8 Current mA ( AC/DC ) measurement
1)Connect BLACK test lead into " COM " terminal ( 3-9,
Fig. 1 ).
2)Connect RED test lead into " mA " terminal ( 3-10, Fig. 1 ).
Open the circuit in which current is to be
measured. Now securely connect test leads in
series with the load in which the current is be
measured.
3)Select the " Function switch " ( 3-7, Fig. 1 ) to the
" mA " position.
4)Power On the meter by pressing the " Power On button "
( 3-2, Fig. 1 ) once, the Display will show " SMART ",
the meter is under " Smart " mode for AC mA, DC mA
measurement.
5)The meter can measure the AC mA, DC mA value
automatically and with auto range selection ( 60 mA,
600 mA ).
5-9 Current 10A ( AC/DC ) measurement
1)Connect BLACK test lead into " COM " terminal ( 3-9,
Fig. 1 ).
2)Connect RED test lead into " 10 A " terminal ( 3-11, Fig. 1 ).
Open the circuit in which current is to be
measured. Now securely connect test leads in
series with the load in which the current is be
measured.
The max. measurement
current for the mA range should
less than 600 mA, otherwise
the protection fuse will be broken.