Astrea Evolve 사용자 설명서 - 페이지 13
{카테고리_이름} Astrea Evolve에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Astrea Evolve 19 페이지. 350mm column instructions for replacing the wetted flowpath using the column refresh kit
STAGE 3 – Refresh the Fixed End
Assembly
Step 1-Disassembly of the Fixed End Assembly.
1.
To assist it is recommended that column feet are lowered to provide a stable base.
2.
Identify the components on the schematic shown in Figure 13.
Figure 13 Fixed Assembly Evolve
3.
Using the correct size allen key, remove and retain the pipespool clamp top.
4.
Remove the pipespool elbow from the flowcell assembly.
5.
Undo, remove and retain bottom flowcell bolts.
6.
Remove the bottom flowcell from the backing plate and discard.
Wetted Flowpath components
DISCARDED
Sanitary Clamp Adaptor
Flowtube seal
Flowtube
Bottom Bed Support assembly
Bottom Flowcell
Pipespool elbow
Bottom Column Flange
Santoprene Ladish Seal
®
Evolve
350mm Column: Instructions for Replacing the Wetted Flowpath using the Column Refresh Kit
®
350 column
ADJUSTER ASSEMBLY
Non-wetted components
RETAINED
Bottom Cell fixing screw
Pipespool clamp and fixing bolts
Backing Plate
Column Levelling feet
Casters
13