Zero Operation
1. Place item to be zeroed on the scale platform.
2. Press ZERO. The scale will display a zero weight. If the item is removed, the
stable negative weight will be shown.
3. The patient and the item are then place on the scale platform. The patient's
weight, less the item's weight, will be displayed.
4. To reset the scale weight to zero, Press Zero again with the scale platform
empty.
NOTE: Stable weights between -5 lb and +5 lb will be zeroed out after 3 seconds.
BMI Operation
1. With a hold weight displayed, press and hold the TARE button for 1.5 seconds
to enter the height value. Height entry starts at 5: 6.0 or 168.0, representing 5
foot 6 inches or 168 centimeters. NOTE: Height units can be changed by
pressing the Units button.
2. Press the ▲ button to increase or the ▼ button to decrease the value. When
the correct value is displayed, press the BMI button to enter value and exit.
NOTE: A menu timer will exit out of this entry mode if no buttons are pressed
within a 30 second time period.
Reweigh Operation
1. With the patient standing on the scale platform, press Reweigh to reinitiate the
weighing process.
2. The display will scroll dashes for about 3-5 seconds until the patient's weight is
captured.
3. The indicator will beep (when enabled) and hold the weight on the display until
the scale shuts off automatically.
8