Comco AccuFlo AF10-E 장비 매뉴얼 - 페이지 9

{카테고리_이름} Comco AccuFlo AF10-E에 대한 장비 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Comco AccuFlo AF10-E 14 페이지. Micro-abrasive blaster

For easier maintenance there are two counters on 
the front panel.
The CYC counts the number of blast cycles on the 
AccuFlo (each time the PowderGate® valve opens.)
Abrasive type and blast pressure will impact 
maintenance frequency, but typical configurations 
will last more than 250,000 cycles before needing 
maintenance.
The HRS counter records the number of blast hours 
(i.e. actual time the modulator has operated.)
Your maintenance requirements may vary with your 
application.
The modulator should be replaced after 5000 hours 
of operation.
To replace the AccuFlo® AF10‐E modulator:
• Disconnect 3 colored tubes from the modulator 
and the electrical plug from front panel.  
• Remove the two screws holding the modulator 
to the bracket using a Phillips screwdriver.
• Install the new modulator in reverse order, 
making sure the two grommets remain in the 
bracket.
Maintenance Counters
AccuFlo® Modulator

MODULATOR

MB2250‐1 (115V)
MB2250‐2 (230V)
Tech Support: 1‐818‐841‐5500
9