filter is compensated for – in other words, the output HV should be independent of load
that the output of the filter, at the J2 or J5 connector, cannot be measured with a conventional DVM 
(input impedance 10 M).  A high impedance probe is required. 
5.3
Thermoelectric Cooler Discussion 
The  control  circuit  for  the  thermoelectric  cooler  is  shown  in  the  schematic  below.    This  circuit 
maintains a fixed  temperature rather than a fixed output voltage.  A temperature sensor is located in 
the detector hybrid.  The conditioning circuitry generates an output of 10 mV/K (2.2 V at 220 K).  This is 
compared to the set point to generate the error signal. 
Figure 7. Schematic for thermoelectric cooler supply circuit. 
For use with a DP5, VR3 is not installed and the DAC controls the value.  For use without a DP5, VR3 
is  installed  (along  with  other  components)  and  the  DAC  is  disconnected.    There  exists  a  maximum 
temperature differential which a given cooler can achieve.  If the set point is lower than this, then the 
supply will provide maximal cooling  (with an output current limit)  but  the detector temperature is no 
longer regulated.  It will change as the system temperature changes.  
A typical two‐stage cooler can generate an 85 °C ∆T. Example: If room temperature is 20 °C, then 
the  maximum  cold  temperature  is  ‐65  °C.  It  is  vital  that  the  detector  temperature  remain  constant. 
Therefore  the  set  point  must  be  set  such  that  no  matter  what  the  ambient  temperature  reaches,  the 
detector temperature stay constant. Example: If room temperature is 40 °C the temperature set point 
must be set such that this is regulated, so the coldest the detector can be set to is ‐45 °C. This means 
that even at 20 °C room temperature the detector will be at ‐45 °C, but it will be constant from 20 to 40 
°C. 
The temperature sensor, inside the detector, is a 1N914.  The forward voltage, at a fixed current, is 
a function of temperature and is used to regulate the detector temperature.  A typical calibration curve 
can be found at www.amptek.com. 
 
                                                             
1
  Note  that  PC5s  earlier  than  Rev  D0  (early  2013)  take  the  feedback  from  before  the  output  filter,  so  that 
there will be some voltage drop across the 2.5Mfilterdepending on current. 
 
Your
Products for
Supply
TEMPMON
TP17
Error
TP16
Amplifier
TEMPSET
 
Imagination
TEC
RC Filter
Power
Temperature
Conditioning
3V
VR3
DAC
11 
1
. Also note 
Sensor