The PUGGLE® Enteral Feeding Pump
the AC adapter provided with the pump.
The PUGGLE Enteral Feeding Pump may be fixed onto a vertical pole using the pole clamp provided with the
pump.
Important:
For the PUGGLE Enteral Feeding Pump and its alarms to operate properly, use only PUGGLE
Enteral Feeding Sets with this device. The pump is not compatible with other pump sets.
Feeding Bag Placement
Place feeding bag above the pump.
For better accuracy, the initial fluid level shall be at
15.2 cm (6.0 inches) higher than the pump and the
pump shall be placed higher than the injection site.
Do NOT use homemade or non-prescribed formula.
The setting range of operating head height is
0~50 cm (0~19.68 inches) which has been
verified as worst-case condition.
1. Open pump cover.
2. Place the silicone loop around the rotor and make sure the cassette is
in the right direction.
3. Push the cassette in the middle of the pins with your thumb and make
sure the cassette is positioned correctly between the two black pins.
Guide both tubes down, in line with their respective slots.
4. Close the pump cover, then place the adapter into the designated slot.
Remove the protective cap from the distal connector (end of down-
stream tube that goes on the patient's feeding tube).
Launch automatic priming.
Connect the tubing to the patient enteral feeding tube and program
the pump.
Scan the QR code to view the
PUGGLE Enteral Feeding Pump
Instructional Video
is delivered with an uncharged battery. Before use, charge by using
Loading Disposable Set
Quick Set Up
Tubing Diagram
1.
2.
Silicone loop
Cassette
Upstream tube
Downstream tube
Adapter
3.