BD LSRFortessa 사용자 지침

{카테고리_이름} BD LSRFortessa에 대한 사용자 지침을 온라인으로 검색하거나 PDF를 다운로드하세요. BD LSRFortessa 2 페이지. Analyzer
BD LSRFortessa에 대해서도 마찬가지입니다: 사용자 설명서 (20 페이지), 지침 (2 페이지)

BD LSRFortessa 사용자 지침
flowkon
Flow Cytometry User Instructions
at the University of Konstanz
FACSFortessa
The BD™ High Throughput Sampler (HTS) option for the BD LSRFortessa™ cell analyzer
provides rapid, fully automated sample acquisition from microtiter plates. In high-
throughput mode, the HTS can process a 96-well plate in fewer than 15 minutes.
Standard throughput mode can be selected for acquisition of larger sample volumes.
Preparing the HTS
Switch the acquisition control to plate mode.
This will switch off the DCM pump.
Unscrew the tube retainer (grey ring
DCM sleeve onto the SIT and carefully remove the sleeve.
Remove the DCM sleeve and replace it with the shorter SIT
protector.
Install the sample coupler on the SIT.
Check the level in the connected water bottle (> ~30 mL).
Switch on the HTS reader (on the backside).
Press the Run button (the HTS does not operate in STNDBY mode).
Initialisation and Priming the HTS
Note: Don't change PMTV of fluorescence channels after enabling compensation.
Operation of the loader is only possible via DIVA software – do not use the prime
button on the instrument.
When the BD DIVA software is already open, go HTS
Start BD DIVA software. The HTS will initialize automatically, otherwise go to HTS
Re-Initialize
Prime the HTS (Sheath tubing connected): HTS
Run your Experiment
See detailed instructions on HTS Programming.
– HTS User Instructions
TM
) that holds the
Re-Initialize.
Prime. 3x in total.