AND TM2440-EX 사용 설명서 - 페이지 2
{카테고리_이름} AND TM2440-EX에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. AND TM2440-EX 2 페이지. Ambulatory blood pressure monitor
Instruction Manual Page 40 8.3.1. A-BPM Items and Parameters
The following changes have been made.
#3: Example for automated measurement.
Start time :
Stores a time. (0 to 23 hour)
Operation time : Set to "OFF"
Response :
A-BPM starts the blood pressure measurement
at the preset Start time and continues until
A-BPM is suspended.
If the EVENT switch is pressing and holding
until the
mark is shown, A-BPM starts at
the preset Start time.
#4: Example for automated measurement.
Start time :
Set to "OFF"
Operation time : Stores time to be continued.
(1 to 27 hours)
Response :
A-BPM starts blood pressure measurement
and stops after the Operation time.
If the
mark is hidden by pressing and
holding the EVENT switch during the
Operation time, A-BPM stops.
If the
mark is shown by pressing and
holding the EVENT switch again, A-BPM is
performed for the Operation time.
Instruction Manual Page 42 8.3.2. A-BPM Program Examples
The following notes were added to
NB. Automatic blood pressure measurement will continue even if turned "OFF" and "ON" again during
automatic measurement.
The following notes were added to "Example 3 Automatic measurement"
NB. Automatic blood pressure measurement will not be performed once the Operation time has been
exceeded even if turned "OFF" and "ON" again during automatic measurement.
Wrong
Example 2 Automatic measurement
"
#3: If the Start time for automatic measurement is specified
and the Operation time is set to "OFF", long pressing
the EVENT switch will start automatic blood pressure
measurement from the Start time set and measurement
will continue until the EVENT switch is long pressed.
If the EVENT switch is long pressed again, automatic
measurement will continue.
If the Operation time is set, automatic blood pressure
measurement will be performed from when the EVENT
switch is initially long pressed until the Operation time
has passed even if the EVENT switch is long pressed
midway through automatic measurement.
#4: If the Start time for automatic measurement is set to
"OFF" and the Operation time is set, long pressing the
EVENT switch will implement automatic blood pressure
measurement after initial automatic measurement until
the set Operation time has passed. If the
switch is long pressed midway through automatic blood
pressure measurement, automatic measurement will
stop. If the
automatic measurement will be performed until the set
Operation time has passed.
If the Start time is set, long pressing the
switch midway through automatic blood pressure
measurement will stop automatic measurement and long
pressing it again will resume measurement.
Addition
Addition
2
Correct
Note
EVENT
switch is long pressed again,
Note
"
1WMPD4004497
EVENT
EVENT