2045Tech FLOOME 사용 설명서

{카테고리_이름} 2045Tech FLOOME에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. 2045Tech FLOOME 1 페이지.

2045Tech FLOOME 사용 설명서
Instructions manual
MOST COMMON REMEDIES
Should the device present any operating anomalies, it is possible to use this table as a reference to solve most common
problems.
PROBLEM
SOLUTION
Device is not recognized by the
Try to turn off and restart the
application
application and try it on another
smartphone.
Be
sure
the
application has access rights to
the microphone
Follow the instructions on the
displayed on the application
application to improve the use of
the device.
Despite
having
followed
the
Verify that no foreign elements are
tutorial, the measurement cannot
present in the whistle chamber. To
be
completed.
The
device
this purpose, use a cotton swab
produces a different whistle from
and inspect the chamber with a
the usual one.
light.
Should the device present further problems, stop using it immediately, and ask for assistance to 2045Tech S.r.l.
SERVICE AND CALIBRATION
is always in optimal conditions, and can deliver you accurate measurements, we advice you to have it calibrated after
app that will advice you to go to the website www.floome.com
to buy the recalibration service
2. The service is very straightforward: a recalibration voucher will be generated, which will have printed on it the address of the
nearest recalibration facility, and a unique code (which will be sent to you by e-mail)
3. Once you have printed the voucher, you will need to send the device through ordinary mail or express courier using the
material while packing the device to avoid any damage during shipping. Thus, we advice you to choose wisely the shipping
method, and consider to opt for a shipping insurance.
will be returned to you by ordinary or express shipping.
SAFETY INSTRUCTIONS
Use the product only for its intended purpose and in accordance with the instructions for use, which can be viewed by visiting
the website: http://www.floome.com - This device is only intended to be used for the detection of blood-alcohol content and not
to determine the capability of an individual to drive also because alcohol consumption is perceived differently from person to
person. However, it is known that the assumption of alcohol compromises the reflexes and the ability to drive exposing the driver
and passengers to potential dangers - This device is intended for private use only and it is not designed and/or intended for
commercial or industrial use. Use the device only if in perfect condition. If any parts are damaged or there is a suspected
Never put the device into water or any other liquids, the device is neither waterproof nor water-resistant. To clean the device, use
PARTS IDENTIFICATION
enumerated parts:
(1) main body of the device, (2) silicone mouthpiece, (3)
circular room, (4) fluid outlet and (5) 3.5mm jack interface.
MAINTENANCE
Repair or maintenance of the device can be carried out only
by experts appointed by 2045Tech S.r.l. Only original parts
may be used for repair. The device contains no part that
will void the warranty. It is required for the user to respect
the intervals, and maintenance methods discussed later in
this manual.
ACCESSORIES AND SPARE PARTS
Only accessories and spare parts explicitly endorsed by 2045Tech S.r.l. are available to the end user for direct application to or
replacement in the product.
INTENDED USE
understood as merely indicative, and not as an evaluation of the state of a person or of his ability to drive.
STORAGE
The device must be kept in an environment, which is compatible with its characteristics. It can be stored without being used for
up to 10 years, but this would compromise the performance of the sensor, that is no longer guaranteed at that point.
FURTHER ADVICE
CLEANING
The device should never be submerged in liquids. The mouthpiece is the only washable part. It can be cleaned only with neutral
If no improvement is detected,
verify
that
the
3.5mm
jack
soap, and has to be reinstalled only once completely dry.
contacts are cleaned.
CE MARKING
T
If necessary, rely on the interactive
handled by 2045Tech S.r.l. - Via delle Cape 3/5 - 30021 Caorle (VE), Italy. For a copy of the declaration of conformity,
tutorial provided in the application.
please visit our website at: http://www.floome.com.
If foreign elements are not easily
FCC MARKING
removable, remove the mouthpie-
T
ce and blow in the air outlet to eject
may not cause harmful interference, and (2) this device must accept any interference received, including interference
the elements from the intake.
that may cause undesired operation.
DISPOSAL OF PRODUCTS
This symbol indicates that this product must not be disposed of with household products. Correct disposal of this product will
help prevent potential negative consequences for the environment, and human health, which could result from improper
handling of the product as waste. For more detailed information about the recycling of this product, please contact your local
MEASUREMENT
measurement process absolutely intuitive.
REQUIREMENTS
(8.1+) operating system. Through the several application marketplaces, which are different depending on the operating
replaces the present measurement instructions.
INSTALLING/REPLACING THE MOUTHPIECE
The device is provided with a mouthpiece, already installed. To replace it, temporarily or permanently, you can simply rotate it
180 degrees, and pull it vertically, until detachment. The new mouthpiece can be installed by simply pushing it onto the
2
protruded part the previous mouthpiece was attached to.
1
REQUIREMENTS OF THE SUBJECT TAKING THE TEST
It is essential for the accuracy of the measurement that the subject has not ingested alcoholic beverages for at least 20
4
3
minutes prior to the test. Residual presence of alcohol in the mouth can affect the measurement, it is therefore necessary to
rinse with water. Ingesting aromatic drinks such as fruit juices, although not alcoholic, can affect the measurement if ingested
overestimation of the measurement. Before testing, the subject has to breathe normally, and just before blowing, he has to be
TAKING THE TEST
5
1. Turn the mouthpiece and extend the 3.5mm jack from the device
4. Follow the instructions in the application to enable the
5. Blow consistently but not too hard up until the smartphone
vibrates or flashes indicating the end of the testing process
7. If the presence of an error is mentioned, try to modify the intensity
or duration of the blow, as suggested by the application, to obtain a
measurement
TECHNICAL SPECIFICATIONS
SIZE
PRODUCER
This device is designed and manufactured in Italy by:
any communication can be made through the website www.2045tech.com or by email to [email protected].
INTELLECTUAL PROPERTY
The design of the device is protected by a U.S. design patent (US D706149 S1) and the vortex flow-meter by a pending
MISCELLANEOUS
causes that may alter the results of the tests, etc.
is a non-registered trademark used by 2045Tech S.r.l.
resistant, in principle. The mouthpiece is made of a non-toxic solvent-free silicone
rubber, suitable for lip contact.
Electrochemical fuel-cell sensor, combined with a vortex whistle flow-meter.
The display interval matches the measurement interval. If a sample with more than
the maximum measurable concentration is introduced, the text "> .250 g/dL" will be
shown.
Operating: 10-40°C, non-condensing humidity between 10 and 90%, ambient
The minimum sampled volume is 0.5L and it requires a duration of the blow of at
least 2.5 seconds.
The device does not require any external power source. It is equipped, inside, with a
2045Tech S.r.l.
Via delle Cape 3/5 - 30021 Caorle (VE)
[email protected] - www.floome.com
Rev. 1.1 - 30/03/2015