Druck InstruMetrics ADTS542F 안전 및 설치 매뉴얼 - 페이지 4

{카테고리_이름} Druck InstruMetrics ADTS542F에 대한 안전 및 설치 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Druck InstruMetrics ADTS542F 16 페이지. Air data test set

1.2 Intended Use
Test and calibration of aircraft aeronautical pressure instruments located indoor in a hanger or
calibration laboratory or outdoor (non-continuously) on a runway, refer to User Manual K0553 for
further information.
WARNING
atmospheres. Using this equipment in a potentially explosive atmosphere could
lead to serious injury or death.
If the equipment is used in a manner not specified by this document and K0553 the protection
provided by the equipment may be impaired.
1.3 Safety
Examine all equipment to be used, paying particular attention to electrical leads and connectors
and pneumatic pipes and connectors. DO NOT use any equipment that is known to be or
suspected to be damaged or faulty.
1.4 Power Supply
Single phase
1.5 Power Supply Connection
Do not apply electrical power outside the values stated.
Make sure that the power supply is off before connecting the power cable.
The ADTS must be connected to the correct electrical power supply as stated, adjacent to the
power connector.
RISK OF ELECTRIC SHOCK
if there is not a protective earth in place. The ADTS must, at all times, be
connected to the supply protective earth/ground.
The power supply cable and connector must be correctly rated for the power
supply. Only use power cables supplied by Druck specifically for the ADTS.
European Color
Green/Yellow
1.6 Fuses
The two High Breaking Capacity fuses (8) (see Figure 6), located in the holders and mounted on
the front panel, protect the unit. The fuses are connected in the live and neutral supply circuit and
are rated at: T5AH 250V.
1.7 Replacing a Fuse
1.
Disconnect the electrical power supply.
2.
Unscrew the fuse holder cap from the front panel and remove the fuse.
3.
Fit a new fuse of the correct size and rating into the fuse holder cap.
4.
Secure the fuse holder cap in the front panel.
Copyright 2014 Baker Hughes Company.
2 | ADTS Safety and Installation Guide–English
This equipment is not rated for use in potentially explosive
110/230 Vac, 50/60Hz
115 Vac, 400Hz
US Color
Brown
Blue
200 VA Max - ADTS542F/ADTS552F
300 VA Max - ADTS553F/ADTS554F
There is a risk of electric shock/electrocution
Black
White
Green
Protective Earth (Ground)
Function
Live
Neutral