the front panel label and are always in the 'ON' state. Tighten each nut a ¼ turn in
succession, trying each corresponding button as you go.
Once each control panel button clicks when pressed, installation is complete.
7.9. Removing the motor
Remove the cabinet assembly by following 7.8
Cut and remove all wire harness zip ties.
Flip the base assembly up-side down and remove the three exhaust air channel
screws with a #2 Phillips driver.
Set the exhaust cover aside
The guard bowl is held in place with six #8 screws. Remove them with a #2 Phillips
driver.
Lift the base assembly off the guard bowl and set it aside.
The motor assembly is held in place with 3 #8 Nylok nuts and fender washers.
Remove the nuts with an 11/32" nut driver.
Remove the three fender washers.
Lift the motor assembly up and off the studs.
The motor is secured to the mounting plate with 4 #8 screws. Use a # 2 Philips
driver to remove the screws.
7.10. Replacing the motor
Fig 1
Place the motor onto the motor mount plate and thread the motor wires into the
cutout as shown (Fig 1).
Attach the motor to the motor mount plate with four #8 screws and torque to 18
in-lbs (Fig 2).
Secure the motor wire harness with a zip tie as shown. Snip the tag end of the zip
tie flush when complete.
DRUCKER DIAGNOSTICS
HORIZON 12 FLEX SERVICE MANUAL
Fig 2
SM028
REV A
7