the front panel label and are always in the 'ON' state. Tighten each nut a ¼ turn in
succession, trying each corresponding button as you go.
•
Once each control panel button clicks when pressed, installation is complete.
7.9. Removing the motor
•
Remove the cabinet assembly by following 7.8
•
Cut and remove all wire harness zip ties.
•
Flip the base assembly up-side down and remove the three exhaust air channel
screws with a #2 Phillips driver.
•
Set the exhaust cover aside
•
The guard bowl is held in place with six #8 screws. Remove them with a #2 Phillips
driver.
•
Lift the base assembly off the guard bowl and set it aside.
•
The motor assembly is held in place with 3 #8 Nylok nuts and fender washers.
Remove the nuts with an 11/32" nut driver.
•
Remove the three fender washers.
•
Lift the motor assembly up and off the studs.
•
The motor is secured to the mounting plate with 4 #8 screws. Use a # 2 Philips
driver to remove the screws.
7.10. Replacing the motor
Fig 1
•
Place the motor onto the motor mount plate and thread the motor wires into the
cutout as shown (Fig 1).
•
Attach the motor to the motor mount plate with four #8 screws and torque to 18
in-lbs (Fig 2).
•
Secure the motor wire harness with a zip tie as shown. Snip the tag end of the zip
tie flush when complete.
DRUCKER DIAGNOSTICS
HORIZON 12 FLEX SERVICE MANUAL
Fig 2
SM028
REV A