Apogee S2-442 소유자 매뉴얼 - 페이지 11

{카테고리_이름} Apogee S2-442에 대한 소유자 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Apogee S2-442 19 페이지. Par-far sensor

Apogee S2-442 소유자 매뉴얼

OPERATION AND MEASUREMENT

Connect the sensor to a measurement device (meter, datalogger, controller) capable of measuring and displaying
or recording a millivolt signal (an input measurement range of approximately 0-150 mV is required to cover the
entire range of illuminance from the sun). In order to maximize measurement resolution and signal-to-noise ratio,
the input range of the measurement device should closely match the output range of the quantum sensor. DO
NOT connect the sensor to a power source. The sensor is self-powered and applying voltage will damage the
sensor.
Wiring for S2-141
Yellow: Positive (signal from PAR sensor)
Blue: Negative (signal from PAR sensor)
White: Positive (signal from Far-red sensor)
Black: Negative (signal from Far-red sensor)
Clear: Shield