Faults
Fault
The scale has switched off
automatically.
The display
shows nothing.
The scale is not connected to the
mains.
Draughts or air movements
Weight display is
Vibrations
unstable.
Electromagnetic fields / static
charging
The scale is not levelled
properly.
The weighing
The scale is not operating in the
result is
specified temperature range.
obviously wrong.
Electromagnetic fields / static
charging
Zeroing is not possible if the
Zeroing is not
displayed weight exceeds +/- 2%
possible.
of the capacity of the scale.
Error messages
Failure
Current weight exceeds the zero
ERR 4
range of the scale
ERR 5
Keypad fault
Load cell is defective or not
ERR 6
connected.
Weighing result unstable due to
draughts, air movement or
Err 9
vibrations.
Load cell is defective.
Current weight exceeds the
Err 19
permissible range for automatic
zeroing when switched on.
Zero range or weighing capacity
--oL--
exceeded
-- UL --
Underload (-20d)
LW100-210408-Rev002-UM-en
Cause
Cause
Measure
Switch on the scale
Use the optional mains adapter to
connect the scale to the mains.
Close the window.
Choose a different location.
Avoid vibrations.
Choose a different location.
Choose a different location. If
possible, switch off the interfering
device.
Adjust the support feet. Use the
spirit level to check the horizontal
position.
Operate the scale only in the
permissible temperature ranges.
Choose a different location. If
possible, switch off the interfering
device.
In this case, switch the scale off and
on again.
Measure
Unload the weighing surface and try
again.
If necessary, restart the scale.
Please contact your service partner.
Please contact your service partner.
Avoid draughts, air movement and
vibrations.
Please contact your service partner.
Unload the weighing surface and
then restart the scale if necessary.
Unload the scale.
Switch the scale off and on again.
13