To set the parameters, go to a saved menu. The parameters are divided into two sections:
one section for the user and the other one for the authorised technician.
To set the parameter menus, press the TARE button during the test run of the display after
switching it on.
The display now shows "P0 CAL" for external calibration.
This function is only reserved for use by authorised technicians!
To navigate through these functions, press the % button to move to the next parameter or
the MODE button to access the previous parameter.
To make changes to the parameter settings, enter the menu by pressing the ZERO button.
To make changes to individual parameter settings, press the % button (moves to the next
higher setting) or the MODE button (moves to the next lower setting).
Press the ZERO button again to save your entries.
Press the G/N button to exit the menu without saving.
Please note that some functions are only reserved for use by authorised
technicians according to the overview below!
PARA-
SUB-PARA-
METER
METER
P0 CAL
P1 LCA
P2 CAP
P3 Fdn
P4 Unt
UNIT
P5 ACC
Accumulation
P6 Ato
Auto Accum.
ZW60-210531-Rev006-UM-en
7. Parameters
DISPLAY
SETTING
External calibration.
Reserved for use by authorised technicians only.
Linearity calibration.
Reserved for use by authorised technicians only.
Capacity setting.
Reserved for use by authorised technicians only.
Division setting.
Reserved for use by authorised technicians only.
Unit setting: kg, g, lb, oz.
on
Selected units enabled.
of
Selected units disabled.
Sum function setting.
on
Sum function enabled.
of
Sum function disabled.
Automatic sum function setting.
Options: Au on / Au off / P Cont / ASK
Au on
Auto on: automatic sum function enabled.
Au of
Auto off: automatic sum function disabled.
P cont
11