AUSTIN SCIENTIFIC M600-HV 매뉴얼 - 페이지 6

{카테고리_이름} AUSTIN SCIENTIFIC M600-HV에 대한 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. AUSTIN SCIENTIFIC M600-HV 13 페이지.

RETURN HELIUM PRESSURE
SUPPLY HELIUM PRESSURE
COOLANT DATA
WEIGHT

3.2 OPERATING AND COOLANT CONDITIONS

When the compressor is operating, the ambient temperature must
remain between 40°F and 104°F (5°C and 40°C).
The standard connections on this compressor, for the coolant system,
are 3/8" Swagelok connectors. Connections can be modified to meet
customer requirements.
Heat exchanger manufacturer recommends 1200-840 micron strainer
on water supply and the following water requirements:
FLOW RATE (MIN) GPM
MIN. INLET TEMPERATURE °F(C)
MAX. INLET TEMPERATURE °F(C)

3.2.1 OPERATING CONDITIONS:

3.2.2 COOLANT CONNECTIONS:

3.2.3 COOLANT QUALITY

Electrical conductivity: >50 µS/cm
Iron:
Manganese:
Ammonia:
pH value:
Chloride:
Free carbon acid:
Nitrate:
Sulphate:
Sulphide:
50HZ PSI (bar)
60HZ PSI (bar)
50HZ PSI (bar)
60HZ PSI (bar)
Lb(kg)
<1.5 mg/l
<0.5 mg/l
<2 mg/l
6 – 9
<100 mg/l
<20 mg/l
<100 mg/l
<50 mg/l
0 mg/l
50
50
50
50
280
280
300
300
1.5
1.5
40(5)
40(5)
85(30)
85(30)
230(105)
230(105)