4.2.3 CONNECTING THE COLD HEAD
Connect the flexline from the compressor supply; this is marked
with a green washer, to the inlet of the cold head. Then connect
the flexline from the compressor return; this is marked with a
red washer, to the outlet of the compressor. This can be
changed with customer requirement.
Crossing the connection on the cold head will cause damage to
the internals of the cold head.
With the appropriate cable supplied from Austin Scientific,
complete the connections form the compressor electronic box to
the electrical connection on the cold head. This will supply
power and logic to the cold head.