Active Motif 53130 사용자 설명서 - 페이지 14

{카테고리_이름} Active Motif 53130에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Active Motif 53130 15 페이지. Multi-sample sonicator

Problem/question
Instrument will not
turn on
Sonication run will not
start
Samples do not sonicate
as expected
www.activemotif.com
Possible cause
Recommendation
No power is being sup-
Check that the power cord is securely plugged into both the
plied to the unit
PIXUL instrument and wall plug.
Unit hasn't been
Ensure that both the power switch on the back side and the
properly turned on
main power switch on the front side of the PIXUL instrument
are both turned on.
Coupling Fluid is not
Check to make sure that the Coupling Fluid has started circulat-
circulating
ing
Plate is installed incor-
Ensure well A1 of the plate is placed in the upper left corner
rectly
Samples are overheat-
Monitor the Coupling Fluid temperature and ensure that it
ing
does not exceed 35°C during sonication. Reduce the Pulse and/
or PRF values if necessary.
Samples are not
Ensure that your sample is diluted in the recommended sonica-
diluted in appropriate
tion buffer, as specified in Quick Guides and User Manuals.
sonication buffer
Consider running a positive control sample, available from
Active Motif.
11