ENSURE THE DEVICE IS COMPLETELY DRY BEFORE USING
DO NOT AUTOCLAVE THE DEVICE
DO NOT REPROCESS THIS DEVICE IN AN ULTRASONIC CLEANER
When decontaminating the McGRATH® MAC EMS with the battery unit installed, disinfection or
sterilization claims will only apply to exposed surfaces of the McGRATH® MAC EMS or McGRATH® 3.6V
EMS Battery. When the battery unit has been removed, newly exposed surfaces should be treated as
contaminated.
Dispose of chemicals & detergents in accordance with local regulations and hospital policy.

3.2 Cleaning

Aircraft Medical used Endozime and Cidezyme Enzymatic Medical Device Detergents when validating
the processes below. Seek approval from your Aircraft Medical agent or distributor before using other
detergents. Always follow the detergent manufacturer's instructions.
CLEANING (MANUAL) – NON-IMMERSIVE
McGRATH® MAC EMS with McGRATH® 3.6V EMS Battery fitted
Prepare Ensure cleaning containers, soft brushes and cloths have been decontaminated
and are in good condition. Prepare a container of detergent in accordance with the
detergent manufacturer's instructions.
Clean
Hold the device above the container, and use soft cleaning brushes to remove
soil from all exposed surfaces, taking care not to scratch optical surfaces. Using a
microfiber cloth, wipe all surfaces thoroughly ensuring that detergent solution comes
into contact with all surfaces. Periodically rinse the cloth and repeat. Scrub crevices
and around hinged/mating surfaces with the soft brush.
Rinse
Hold the device over a large volume of rinse water and wipe down to remove all traces
of detergent.
Repeat
The cleaning and rinsing steps should be repeated if the device was heavily soiled or if
visible soiling remains after cleaning.
Dry
Carefully hand-dry using clean, absorbent, non-shedding cloth.