3.2 Cleaning
CLEANING (MANUAL) – NON-IMMERSIVE
McGRATH® 3.6V EMS Battery
Prepare
Ensure cleaning containers, soft brushes and cloths have been decontaminated
and are in good condition. Prepare a container of detergent in accordance with the
detergent manufacturer's instructions.
Clean
Hold the battery unit above the container, and use soft cleaning brushes to remove
soil from all exposed surfaces. Using a microfiber cloth, wipe external surfaces
thoroughly ensuring that detergent solution comes into contact with the surfaces.
Periodically rinse the cloth and repeat.
The battery unit is NOT immersible. Do not saturate the electrical contacts or the
cavities surrounding the retaining clips with detergent or rinse water. If the battery
unit is heavily soiled do not re-use. Discard accordingly.
Rinse
Hold the battery unit over a large volume of rinse water and wipe down to remove
all traces of detergent. Again, take care not to saturate the electrical contacts and
cavities surrounding the retaining clips with water.
Repeat
The cleaning and rinsing steps should be repeated if visible soiling remains after
cleaning.
Dry
Carefully hand-dry using clean, absorbent, non-shedding cloth.
CLEANING (MANUAL) - IMMERSIVE
McGRATH® MAC EMS with McGRATH® 3.6V EMS
Battery removed
Prepare
Ensure cleaning containers or sinks, soft brushes and cloths have been
decontaminated and are in good condition. Prepare a container of detergent in
accordance with the detergent manufacturer's instructions.
Pre Clean If visible soiling is present, immerse the device for a minimum of 5 minutes in
enzymatic detergent solution prior to cleaning.
Clean
Immerse the device for 5 minutes in fresh enzymatic detergent solution. Then,
keeping the device immersed, use soft cleaning brushes to remove additional soil
from challenging design features. Take care not to scratch optical surfaces. Scrub
crevices and around hinged/mating surfaces with a soft brush.
Rinse
Remove the device from the solution and drain over the detergent solution before
transferring to a clean-rinse receptacle or sink. Rinse device thoroughly with clean
water, ensuring that it is fully immersed. Remove device from rinse water.
Repeat
The cleaning and rinsing steps should be repeated if the device was heavily soiled or
if visible soiling remains after cleaning.
Dry
Carefully hand-dry using clean, absorbent, non-shedding cloth.