THE PROCEDURE WE DESCRIBED ABOVE PROVIDES THE EXPECTED VALUES FOR A DEVICE
EQUIPPED WITH VACUUM TUBES PROVIDED BY GAINLAB AUDIO.
IF YOU STILL WANT TO USE YOUR EQUIPMENT WITH 3RD PARTY TUBES, MAKE SURE THAT THE
TECHNICAL PARAMETERS OF THE VACUUM TUBES THAT YOU PLACED IN THE TUBE1, TUBE2 SOCKET
AND THE TUBE3, TUBE4 SOCKET ARE APPROXIMATELY THE SAME. MATCHED PAIR TUBES ARE
SUGGESTED TO USE.
IF YOU DO NOT USE VACUUM TUBES WITH THE SAME TECHNICAL PARAMETERS, THE CALIBRATION
PROCESS WILL NOT BE PERFORMED CORRECTLY!
THE TECHNICAL PARAMETERS OF THE VACUUM TUBES CAN CHANGE SIGNIFICANTLY IN THE FIRST
24-48 HOURS. IT IS RECOMMENDED TO REPEAT THE CALIBRATION PROCESS AFTER THIS "BURN-
IN" PERIOD.
THE VACUUM TUBES PROVIDED BY GAINLAB AUDIO HAVE ALREADY UNDERGONE THIS "BURN-IN"
PROCESS.