5.7 Troubleshoot
LED indications
Situation
Description
Connected via USB Green continuous
Orange blinking
Multicolor blinking
Red continuous
During
Green blinking
measurement
Green continuous
Orange blinking
Error indications
2 rapid red blinks, pause repeated 3 times in total
3 rapid red blinks, pause repeated 3 times in total
4 rapid red blinks, pause repeated 3 times in total
5 rapid red blinks, pause repeated 3 times in total
6 rapid red blinks, pause repeated 3 times in total
Troubleshoot for error indications
Error Type*
Troubleshoot
Error 1
Disconnect and replug the Physilog®
Error 2
Contact Gait Up
Error 3
Charge the Physilog®
Error 4
Contact Gait Up
Error 5
Contact Gait Up
Error 6
Empty the memory card
*see LED indication above
Soft reset
The soft reset is the first manipulation which can be done if the Physilog® is not working
properly. To do a soft reset, hold the main button for 15 seconds. The LED shines orange and
then stops. If the problem persists, fill out the contact form on the support page
(www.gaitup.com/support2017) or write an email to: [email protected]. Please indicate
Physilog® serial number and describe the problem in as much detail as possible.