CLEANING INSTRUCTIONS
CUSHION
1. Use germicidal soap to wash the cushion.
2. Use a wash cloth to wipe the cushion dry. Do not use detergents
or other solvents.
3. Do not machine wash or tumble dry.
4. Do not heat or steam autoclave.
COVER
1. Wash at 30˚C / 86˚F
2. Do not iron.
3. For single patient use only.
REPAIRING LEAKS
1. Make sure the cushion and all four corners are inflated
completely.
2. Immerse the over-inflated cushion into water to locate the leak.
(You will see air bubbles rising from the area where the leak
is located.)
3. When the leak has been located, mark the leak.
4. Deflate the cushion.
5. Allow the cushion to dry completely.
6. Clean the area around the hole and let the surface dry.
7. Place a patch film over the hole.
8. Press the patch film down firmly to seal. Do not use the cushion
for at least 30 minutes.
9. When leak repair is complete, adjust the cushion to proper
inflation according to the user manual.
10. Repeat to test if there are any other leaks.
Patches do not change the performance of the cushion, but if your
patch is not successful in solving the problem, or there is more
serious damage, please contact your local provider.