Avantes AvaTrigger-USB2 운영 매뉴얼 - 페이지 2
{카테고리_이름} Avantes AvaTrigger-USB2에 대한 운영 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Avantes AvaTrigger-USB2 2 페이지. External trigger box
AvaTrigger-USB2 External Trigger Box
Operating Manual
The AvaTrigger is a trigger source that can be used to trigger
an AvaSpec-USB2/EVO spectrometer. The AvaTrigger has two
different settings for external triggering, i.e. an optical
trigger and a pushbutton trigger.
The 3-position switch should be set into the left position to
use the optical trigger. If triggers should be generated
manually by the pushbutton, set the switch to the right. The
trigger box is switched off by setting the switch in the
middle position.
The optical trigger is useful to measure pulsed light sources, the AvaSpec can start integration time
within 1.5 microsecond (depending on the model) after the optical signal starts, or a user specified
delay time later if the spectral output against time needs to be measured. The optical trigger has an
SMA905 connector to easily couple to our comprehensive line of light sources and accessories
through SMA couples fiber optics. The sensitivity of the optotrigger can be adjusted by a
potentiometer at the front side of the AvaTrigger. The green LED at the front side shows a short
pulse when a TTL pulse is put at pin 4 of the DB15 connector at the backside of the AvaTrigger box.
The AvaTrigger comes with an IC-DB26-2 interface cable. This cable needs to be connected between
the AvaTrigger and the AvaSpec-USB2/EVO spectrometer.
Technical Data
Trigger Input
Internal Delay time to
TTL output
Power consumption
IO connectors
Dimensions
Weight
*depending on the slew rate of the light source
Revision: 1-00
© Avantes BV
Opto
Ca. 300 ns*
5 mA @ 5VDC (internal)
DB15 at AvaTrigger
3
10
4
75 x 78 x 37 mm
www.avantes.com
Pushbutton
20 µs
DB26 at AvaSpec-USB2
15
1
6
260 g
1