Ceyear 2498D 사용자 설명서 - 페이지 12

{카테고리_이름} Ceyear 2498D에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Ceyear 2498D 15 페이지. Optical power meter

Section II Working Principle of Optical Power Meter

The working principle diagram is shown in Fig. 3-1. The program-controlled amplifier circuit converts the
measured optical power into a photocurrent proportional to it, and then sends it to an analog-to-digital
converter to convert it into a digital quantity representing the magnitude of the optical power, and further
sends it to the control processing section for calculation and processing, and then sends it for displaying.
The control processing part also performs operations such as zero setting, turning on the laser,
measurement mode conversion, wavelength setting, and corresponding calculation according to the key
operation.
Optical input
Optical
output
Chapter III Main Technical Indicators and Working Principles
Optical input
Light source control
circuit
Figure 3-1 Block Diagram of the 2498C/D optical power meter
EEPROM
data memory
STM32
MCU &
analog-digital
converter
Upper cover plate
component
Key
LCD
Power supply
power-up module
USB
Data
communication
module
9